Labman Portable Microprocessor UV Visible Spectrophotometer

Wavelength Range 190 nm
Spectral Bandwidth 1 nm
Mobility Portable
Type Portable
Display 320 x 240 Graphic LCD
Optical System Double Beam, Grating 1200 lines / mm
Brand Labman
Model Name/Number LMSP-UV1900
Photometric Accuracy +0.3%T
Photometric Range -0.3-3A, 0-200%T, 0-9999C
Photometric Repeatability + 0.2%T
Data Output Port USB
Cell Holder 10mm single cuvette Holder
Dimensions 625 x 430 x 206 mm
Printer Port Parallel Port
Wavelength Repeatability + 0.2 nm
Material ABS Plastic
Wavelength Accuracy + 0.3nm
Stray Light <0.05%T @ 220nm, 360nm
Stability + 0.001A/h @ 500nm
Lamps Deuterium Lamp & Tungsten Halogen Lamp
Weight 30 kg
Detector Silicon Photodiode
Bandwidth 1nm
Scanning Speed Fast, Medium & Low
Baseline Flatness & Noise +0.001A
Power Supply AC 220V/50Hz

Features:

  • The innovative optical layout and state of art mono- chromator with high grade blazed holographic grating ensure  accuracy. Its integrated design assures long term stability  and durability
  • The precisely aligned detector and quality deuterium and halogen lamps enhance the precision across the UV / Vis spectrum starting from 190 nm and into the near infrared 1100nm
  • The two detectors are used to measure sample and reference respectively and simultaneously for optimizing  measurement accuracy
  • Lamp Selection  enables  conserving the life of D2 & Tungsten lamps. Real time clock for date & time stamping of results
  • Big LCD display. (320 x 240 Dots Matrix)
  • Basic Mode, Quantitative, Kinetics, Multi – Wavelength, Wavelength Scan, DNA / Protein Test &Utility

Function Of Software:

  • Basic Mode
  • To measure the Absorbance and Transmittance

Quantitative:

  • Coefficient Method
  • Standard Method curve.
  • Up to 10 standard samples may be used to establish a curve. Four methods to fitting a curve through the calibration points: Linear fit through zero, square fit & cubic fit.

Wavelength Scan:

  • 1. The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2 & 5 nm
  • 2. Fast, Medium & Slow Scan speeds are available. They very from100 to     1800 nm/min
  • 3. Wavelength are scanned from high to low so that the instrument waits at     high Wavelength and minimizes the degradation of UV sensitive samples

Multi-Wavelength:

  • Up to 32 wavelengths can be selected and multiple samples can be measured

Kinetics:

  • This mode may be used for time course scanning or reaction rate calculations. Abs. vs. time graphs is displayed on the screen in real time
  • Wait time and measurement time up to 12 hours may be entered with time intervals of 0.5,1,2,5,10,30 seconds & 1min. Post- run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor

DNA/Protein Test:

  • Concentration and DNA purity are quickly and easily calculated: Absorbance ratio s 260nm/280nm with optional subtracted absorbance at 320nm.
  • DNA concentration= 62.9x A260-36.0x A280
  • Protein concentration=1552xA260-757.3xA280

Utility:

  • Has following functions
  • WL Reset, Printer, Lamp, Clock, Dark Current, Accu. Validity, WL Validity Connect to PC, Beeper On/Off & System Base Line

Reviews

There are no reviews yet.

Be the first to review “Labman Portable Microprocessor UV Visible Spectrophotometer”