Analysis BOD Incubators are designed primarily to meet the requirements for tests of Biological Oxygen Demand / Biochemical oxygen demand and equipped for controlled incubation and storage of sensitive samples / cultures, vaccines etc. . These incubators are designed to meet the testing challenge of pharmaceutical competitive research field and enhance the scientific efforts of educational institute and organisation. Analysis Make GMP BOD Incubator is especially built to Pass FDA and ICH Q1A compliance standards and have a reputation for reliability, quality and performance. Exceed FDA/ICH guidelines for long term, intermediate and accelerated stability studies.

BOD Incubator Equipments

Specifications :

Temperature Range 5.0 °C to 60.0 °C. Accuracy +/- 0.1 °C. Uniformity +/- 1.0 °C

 Temperature Display : Digital LED 3 ½ digit, Resolution 0.1 deg C

 Heating system: U shaped Nichrome wire heater in SS Sheathing.

 Cooling system: Hermetically sealed CFC free Emerson compressor with R 134a refrigerant.

 Air circulation: Flange Motor with Impeller and blower.

 Interior Illumination for working Area.

 There will be two nos of blower from 450 Ltr Onwards.

Construction :

Double walled, gap between walls filled with PUF Insulation.

 Standard Model: Stainless steel Mirror polished chamber and heavy gauge Powder- coated CRC exterior.

 GMP Model: Stainless steel Mirror polished chamber and exterior of dull finish stainless steel.

 Double door: Inner full Viewing door and outer door insulated with lock and key arrangement.

 50mm Validation Port with silicone rubber seal to insert sensor for validation purpose.

 3” Thick CFC Free Puff Insulation.

 Stainless steel Trays. Adjustable Tray Height adjustment.

 Unit Mounted on castor wheels with front lock arrangement.

Control Panel :

Microprocessor based PID temp. Controller with auto tune facility for precise control of temperature.

 Temp. Sensor: PT- 100.

 Dedicated safety controller with separate sensor to cut off the supply in case of overshoot and undershoot of temperature giving audio visual alarm. ( Optional)

 Recorder data bank to store and print set values , process values , date and time of master controllers with variable print time intervals ( Optional)

 Alarm: Audio visual alarm if temperature deviates from the present temperature.

 PC Interface communication card with RS 485 for temperature and humidity control system ( Hardware only) ( optional)

Safety Features :

Utilites :

Over heat temperature with alarm.

 Electrical overload compressor cut-off.

 Time delay for compressor switch on.

 Electrical short circuit breaker.

Room temperature around machine preferably at 25 degree C.

 Stabilised input voltage of 230V Ac, 50 Hz, and 20 Amps.

 Water drains line with ½” Nozzle to be provided at floor level.

Data Record :

21 CFR Part 11 compliant software with audit trial ( who did what and when is recorded)

 Multilevel Passwords.

 Data storage device has online and offline capabilities.

 Multi chamber data acquisition on single software.

 Tabular and graphical report, alarm deviation report.

 Mean Kinetic temperature report and alarm deviation report.

 Audit trial report.

 Attractive reports and graphs.

 Alarm report.

 Time synchronisation.